|
|
|
Instructions,
|
Reference: Inhalation System for Pulmonary Aerosol Drug Delivery in Rodents Using Large Porous Particles. Abdellaziz Ben-Jebria, Mary Lou Eskew, and David A. Edwards. Aerosol Sci. and Technol. 21:421-433 (2000). Drug Delivery by Inhalation Using Large Porous Particles in
Rodents
Effective therapies directed against asthma have advanced since 1956 when the
first pressurized metered dose inhaler was developed. The target for these
anti-asthma medications is the smooth muscle cells concentrated in the pulmonary
airways. Delivery of drugs directly to the lungs is an effective and relatively
safe means of both local and systemic treatment of a variety of medical
conditions. Medication delivery may be inefficient if particles aggregate in the
inhaler itself, if they are filtered by the throat and mouth, and if mucociliary
clearance processes remove them. In order to keep dry powder aerosols from
depositing in the oropharyngeal cavity and inhaler, most have a geometric
diameter of 1-3 um and particle mass density of ~ 1g/cm3. Unfortunately
particles of this size also aggregate easily and are then phagocytosed by
pulmonary macrophages resulting in a decrease in their therapeutic potential. Drug development for pulmonary use necessitates inhalation toxicology
studies. Most of these studies require the use of relatively large amounts of
drug due to losses before and during exposure. Ben-Jebria et al. (1999)
developed unique particles for inhalation use. These particles are porous with a
large geometric diameter (~10 um) and low mass density (<0.1g/cm3). They
hypothesized that these particles would be too large to be phagocytosed by
macrophages and would, therefore, remain in the lungs longer than conventional
particles. An experimental model was developed using rats and guinea pigs to
measure the efficacy of drug delivery using these particles. Both polymeric and nonpolymeric excepients were used to prepare these large
particles with low mass density. Solvent evaporation and spray drying was used
in both cases as described in detail in the reference (below). The polymeric
microsphere preparation was made of biodegradable poly(lactic acid-co-glycolic
acid) (PLGA) polymer particles. The nonpolymeric large porous particles
consisted of human serum albumin, lactose, dipalmitoyl phosphatidylcholine (DPPC)
and albuterol sulfate. Particles were also made of DPPC, albumin, lactose, and
estradiol. The DPPC allows these nonpolymeric particles to be relatively water
insoluble permitting a sustained release of hydrophilic or lipophilic drugs such
as albuterol or estradiol over one to two days. The DPPC particles all had
geometric diameters of ~10um and bulk (tapped) densities of approximately 0.08
g/cm3. Control non-porous particles made of lactose and albuterol had a smaller
geometric diameter and larger bulk density (diameter=3u and bulk density of 0.45
g/cm3). These particles were aerosolized with a Spinhaler dry powder inhaler and
characterized in vitro with a cascade impactor and Aerosizer as detailed in the
reference. About 50% of the large porous particles had a MMAD of <4.7 um
while 16% of the small nonporous particles were <4.7 um or "fine
particles". In addition, Spinhaler results showed that large porous
particles were less aggregated than nonporous particles. The MMAD of large
porous particles was ~2.15 um while the MMAD of nonporous particles was ~ 4.53
as measured with the Aerosizer. The smaller MMAD results in greater
respirability in vitro. The final test of the utility of large porous particles for both delivery and
sustained release into the lungs required the development of a novel rodent
exposure system. Rats were used to test the release over time of insulin,
testosterone, and estradiol. Guinea pigs were used to assess the bronchodilation
effect of albuterol. The animal is anaesthetized and placed supine on a surgical
board. The proximal trachea is exposed and a blunt cannula inserted and
connected to a ventilator as shown in Figure 1. The animal is still able to
breathe through proximal and upper airways because the outer diameter of the
cannula is is 1 mm whereas the inner diameter of the trachea is 3-4 mm. (The 1mm
diameter cannula has greater airway resistance than the 3-4 mm airway). A
pipette with a rubber bulb is attached to the inhalation side of the ventilator
tubing for the purpose of introducing the particles. The bulb is filled with a
set amount of particles and a portion is introduced as the operator squeezes it
while the respirator is set to inhale. The portion is delivered in one 30 to 60
sec insufflation by forced ventilation at 3 ml tidal volume and 100 strokes/min
frequency. Several repetitions of the procedure deliver 3 to 5 mg of particles
over a period of 10-15 min. Figure 2: Total particle mass (percent of mass
delivered) recovered from rat respiratory tract after bronchoalveolar
lavage. "Respirable" is the fraction of the delivered particle
mass that is recovered in rat lungs beyond the carina, after bronchoalveolar
lavage; and "Nonrespirable" is the fraction of teh delivered particle
mass that is recovered in the tubing and trachea. (from Ben-Jebria et al,
2000) The bronchoalveolar lavage fluid (BALF) was collected 10 min after the
inhalation procedure and the number of particles in each lobe of the lobe was
counted as described in the reference. Figure 2 shows that most of the large
porous particle mass was deposited in the lungs distal to the carina. Less than
45% was left in the tubing and trachea. However only 20% of the mass of small
nonporous particles was deposited in the respiratory tract. There were no
significant changes in cell population or protein recovered from the BALF in
guinea pigs treated with large porous particles with and without albuterol. The bioavailability of the drugs, insulin, testosterone and estradiol
administered to the lungs with large porous particles was compared to the
bioavailability when administered subcutaneously. Radioimmunoassay was used to
quantitate the amount of insulin, estradiol and testosterone. 87.5% more
estradiol, 177% more insulin, and 86% more testosterone were available from
inhaled large porous particles compared to subcutaneous administration of these
drugs. In contrast the small nonporous particles yielded less bioavailable
insulin (12%) and estradiol (18%). Blood concentrations of these drugs remained
elevated significantly longer when they were administered on large porous
particles compared to administration on nonporous particles. There was also a
sustained inhibition of carbachol-induced bronchoconstriction when guinea pigs
inhaled large porous particles with albuterol for 15 min. The authors conclude that large porous particles reach the deep lung more
efficiently than do more conventional small nonporous particles in this
experimental rodent model. In addition large porous particles have a longer
residence time in the lungs than do small nonporous particles. The large porous
particles have relatively low mass density. This causes better penetration of
the large porous particle because the aerodynamic diameter of these particles is
similar to that of nonporous particles. Greater aerosolization efficiency allows
less chance of loss in large airways and tubing. The insufflation technique for
aerosol delivery described is concluded to be useful for dry powders because it
allows a small amount of aerosol to be delivered in a short period of time. Reference: Inhalation System for Pulmonary Aerosol Drug Delivery in
Rodents Using Large Porous Particles. Abdellaziz Ben-Jebria, Mary Lou Eskew, and
David A. Edwards. Aerosol Sci. and Technol. 21:421-433 (2000). Also: Ben-Jebria, A., Chen, D., Eskew, M.L., Van-bever, R., Langer, R.,
and Edwards, D.A. (1999). Large Porous Particles for Sustained Protection from
Carbachol-induced Bronchoconstriction in Guinea Pigs, Pharm. Res. 16:555-561. By: Susan G. Shami, ScD
|